FDAClass IIMarch 14, 2018

Medtronic Synchromed II recall

Medtronic Synchromed II Model 8637 supplied in 20mL and 40 mL reservoir size and a catheter. Product Usage: The implantable Model 8637 SynchroMed II…

Why was Medtronic Synchromed II recalled?

Medtronic implemented a final design change intended to address the motor stall due to corrosion and is recalling devices with the previous design that are not implanted.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

US: CFN #: GTIN: PIN: 8637-20 00643169508149 863702012H 8637-40 00643169508156 863704012H OUS: CFN#: GTIN : PIN: 8637-20 00643169700901 863700221K 8637-20 00643169700918 863700222K 8637-20 00643169700925 863700223K 8637-20 00643169700932 863700224K 8637-20 00643169700949 863700225K 8637-20 00643169700956 863700226K 8637-20 00643169700963 863700227K 8637-20 00643169700970 863700228K 8637-20 00643169700987 863700229K 8637-20 00643169700994 863700230K 8637-20 00643169732247 863700241K 8637-20 00643169732254 863700242K 8637-20 00643169732261 863700243K 8637-20 00643169732278 863700244K 8637-20 00643169732285 863700245K 8637-20 00643169732292 863700246K 8637-20 00643169732308 863700247K 8637-20 00643169732315 863700248K 8637-20 00643169732322 863700249K 8637-20 00643169732339 863700250K 8637-40 00643169701007 863700231K 8637-40 00643169701014 863700232K 8637-40 00643169701021 863700233K 8637-40 00643169701038 863700234K 8637-40 00643169701045 863700235K 8637-40 00643169701052 863700236K 8637-40 00643169701069 863700237K 8637-40 00643169701076 863700238K 8637-40 00643169701083 863700239K 8637-40 00643169701090 863700240K 8637-40 00643169732346 863700251K 8637-40 00643169732353 863700252K 8637-40 00643169732360 863700253K 8637-40 00643169732377 863700254K 8637-40 00643169732384 863700255K 8637-40 00643169732391 863700256K 8637-40 00643169732407 863700257K 8637-40 00643169732414 863700258K 8637-40 00643169732421 863700259K 8637-40 00643169732438 863700260K

Product
Medtronic Synchromed II Model 8637 supplied in 20mL and 40 mL reservoir size and a catheter. Product Usage: The implantable Model 8637 SynchroMed II…
Recalling firm
Medtronic Neuromodulation
Quantity
27,547 US and 13,025 OUS
Distribution
US Nationwide Distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OR, OK, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY
Recall date
March 14, 2018
Classification
Class II
Status
Ongoing
Agency
FDA (device)