GE Healthcare Discovery NM/CT 670 CZT recall
GE Healthcare Discovery NM/CT 670 CZT, Model/Catalog Numbers: 1) H3906AB; System, Tomography, Computed, Emission
Why was GE Healthcare Discovery NM/CT 670 CZT recalled?
GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
1) H3906AB: UDI/DI 840682121446, System ID UX2305NU03, Serial Number CZTV29005; UDI/DI 840682121446, System ID M4166103, Serial Number CZTV29006; UDI/DI 840682121446, System ID 00243NUC22, Serial Number CZTW29008; UDI/DI 840682121446, System ID 00222NUC07, Serial Number CZTW29009; UDI/DI N/A, System ID 604875D670B, Serial Number CZTW29007; UDI/DI N/A, System ID NT6200, Serial Number AD7W38007; UDI/DI N/A, System ID NT6201, Serial Number AD7X38017; UDI/DI N/A, System ID NT6100, Serial Number AP7W37001.
- Product
- GE Healthcare Discovery NM/CT 670 CZT, Model/Catalog Numbers: 1) H3906AB; System, Tomography, Computed, Emission
- Recalling firm
- GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
- Quantity
- 8 units
- Distribution
- Worldwide
- Recall date
- January 22, 2025
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)