Orthofix ISKD Limb Lengthener. ISKD - Intramedullary recall
Orthofix ISKD Limb Lengthener. ISKD - Intramedullary Skeletal Kinetic Distractor. The ISKD is intended for limb lengthening of the tibia and femur.
Why was Orthofix ISKD Limb Lengthener. ISKD - Intramedullary recalled?
There is a possibility that the ISKD limb lengthener may stop distracting post-operatively during treatment, which may result in premature bone consolidation (limb not achieving the desired length) leading to revision surgery to remove and/or replace the device.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All serial numbers of the following Catalog Numbers: F12-255-305, F12-345-395NS, T12-245-295, F12-255-305NS, F12-345-425, T12-255-305, F12-255-335, F12-345-425NS, T12-255-305NS, F12-255-335NS, T10-215-265, T12-255-335, F12-300-350, T10-255-305, T12-255-335NS, F12-300-350NS, T10-255-335, T12-300-350, F12-300-380, T10-300-350, T12-300-350NS, F12-300-380NS, T12-215-265, T12-300-380, F12-345-395, T12-215-265NS, and T12-300-380NS
- Product
- Orthofix ISKD Limb Lengthener. ISKD - Intramedullary Skeletal Kinetic Distractor. The ISKD is intended for limb lengthening of the tibia and femur.
- Recalling firm
- Orthofix, Inc
- Quantity
- 1224
- Distribution
- Worldwide Distribution - USA (nationwide) Internationally to Australia, Hong Kong, Saudi Arabia, Korea, Turkey, Greece, Finland, South Africa, Spain, Singapore, Switzerland, Brazil, Germany, France, Japan, Netherlands, New Zealand, Israel, Canada and Mexico
- Recall date
- March 6, 2013
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)