FDAClass IMarch 13, 2013

Ad-Tech Medical Instrument Corporation Macro Micro recall

Ad-Tech Medical Instrument Corporation Macro Micro Subdural Electrode. For temporary (<30 days) use with recording, monitoring and stimulation…

Why was Ad-Tech Medical Instrument Corporation Macro Micro recalled?

The microelectrodes are not flush with the silastic substrate, may contain sharp edges capable of abrading the cortical surface, and may have remaining pieces which may fracture and remain in the cortical tissue with the electrode is explanted. Recall expanded on 5/2/2013 to include additional Macro Micro Subdural Electrodes.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

REF numbers (catalog numbers): EG10A-SP10N-000, EG34B-SP10X-000, EGD7A-MP12N-000, EGE4A-SP10N-000, EGE4A-SP10X-000, QG10B-SP10X-000, QG10C-SP10N-000, QG10D-SP10N-000, QG11A-SP10X-000, QG16A-SP10X-000, QG17A-SP10X-000, QG18A-SP10N-000, QG18A-SP10X-000, QG20B-SP10N-000, QG20B-SP10X-000, QG20C-SP10N-000, QG20C-SP10X-000, QG20D-SP10X-200, QG20E-SP10X-000, QG20F-SP10X-000, QG24A-SP10N-000, QG26A-SP10N-000, QG32A-SP10X-000, QG32B-SP10X-000, QG36A-SP10N-000, QG36A-SP10X-000, QG36B-SP10N-000, QG44A-SP10X-000, QG53A-SP10X-000, QG66A-SP85N-000, QG69A-SP10N-000, QG88A-SP10X-000, QG96A-SP10N-000, QG96A-SP10X-000, QG96C-SP10X-000, QG96D-SP10X-000, QGC6A-SP07N-000, QGC8A-SP10N-000, QGC8A-SP10X-000, QGC8B-SP10X-000, QGC8C-SP10X-000, QGC8D-SP10N-000, QGC8E-SP10X-000, QGC8F-SP10X-000, QS20A-SP10N-000, QS40B-SP10X-000, Lot numbers: 208140314, 208140343, 208140344, 208140353, 208140354, 208140362, 208140369, 208140374, 208140381, 208140382, 208140387, 208140390, 208140395, 208140398, 208140399, 208140414, 208140417, 208140421, 208140431, 208140443, 208140445, 208140450, 208140452, 208140453, 208140458, 019020, 208140356R, 208140357R, 208140452R, C010180001, C702201001, C703160006, C704040001, C800250002, C800350004, C801240003, C802220003, C802220005, C802220006, C805210003, C805210004, C908221002, G012111001, G012140002, G015020002, G016240001, G017250004, G017250005, G017250006, G018011001, G018011002, G018011003, G019021002, G019021003, G019150003, G112270004, G114160001, G114160002, G115080002, G115170004, G115280005, G119020004, G211220001, G608060002, G704150001, G805270001, G900211002, G903011001, G903080003, S903080004, S905280003. Recall exanded on 5/2/2013 to include additional Macro Micro Subdural Electrodes: REF numbers (catalog numbers): CG-64BPX-F436, CMMS-22PX-F478, , CG-32BPX-F412, CGIE-98BPX-F341, CGIE-82BPN-F377, CMMS-69PN-F480, CGIE-98BPN-F341, CG-9SN-F496, CG-32BPX-F426, CG-16BPX-F445, CG-16BPX-F446, CG-256PX-F425, CGIE-39BPX-F442, CG-46BPX-F465, CG-64BPX-F419, CG-32BPX-F418. Lot numbers: 20814044, 208140447, 208140463, 208140463, 208140476, 208140463, 08140463, 208140424, 208140439, G119221001,G212040002, G216040002, 218190, 208140431, 208140450, 208140432, 208140433, 208140420, 208140437, 208140450, 208140422, 208140422, 208140422, 208140422

Product
Ad-Tech Medical Instrument Corporation Macro Micro Subdural Electrode. For temporary (<30 days) use with recording, monitoring and stimulation…
Recalling firm
Ad-Tech Medical Instrument Corporation
Quantity
144 (103 US; 41 OUS)
Distribution
Worldwide Distribution - USA Nationwide including the states of CA, GA, LA, MA, MD, MI, MN, NH, PA, RI, TX, UT, WI, NY IA, WI and the countries of: Australia, CANADA, ENGLAND, JAPAN, KOREA, SPAIN, FRANCE, and SWITZERLAND.
Recall date
March 13, 2013
Classification
Class I
Status
Terminated