FDAClass IIDecember 17, 2025

Safe-T-Centesis 6 Fr Catheter Drainage Tray recall

Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,

Why was Safe-T-Centesis 6 Fr Catheter Drainage Tray recalled?

Acute drainage catheters, indicated for use in removing fluid during acute drainage/aspiration procedure, manufactured with safety indicators that erroneously show catheter in sharp position when catheter is in blunt position, which may lead users to believe device is not yet in pleural or peritoneal space when it is, may cause discomfort, delay to treatment, injury to internal organs or vessels.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI/Lot/Expiration: (01)10885403108365(17)271130(10)0001623320/0001623320/20271130; (01)10885403108365(17)271130 (10)0001623069/0001623069/20271130

Product
Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,
Recalling firm
Bard Peripheral Vascular Inc
Quantity
1240
Distribution
US Nationwide distribution in the states of CA, MN, IN, MI, LA.
Recall date
December 17, 2025
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls