High Speed Cell Saver¿5/5+ Bowl Set Autotransfusion recall
High Speed Cell Saver¿5/5+ Bowl Set Autotransfusion Device - Product Usage: its related accessory components are intended to be used for the recovery…
Why was High Speed Cell Saver¿5/5+ Bowl Set Autotransfusion recalled?
Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Catalog number/FG Item Number: 0263A-00 1. No lots for cracking issue identified (0 Distributed). 2. Missing instructions to check for cracks in IFU All products/lost manufactured between October 2014 and October 2019.
- Product
- High Speed Cell Saver¿5/5+ Bowl Set Autotransfusion Device - Product Usage: its related accessory components are intended to be used for the recovery…
- Recalling firm
- Haemonetics Corporation
- Quantity
- 3563
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
- Recall date
- February 5, 2020
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)