SIGNA Pioneer Nuclear Magnetic Resonance Imaging System recall
SIGNA Pioneer Nuclear Magnetic Resonance Imaging System - Product Usage: is a whole body magnetic resonance scanner designed to support high…
Why was SIGNA Pioneer Nuclear Magnetic Resonance Imaging System recalled?
It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system's date and time setting are used to populate the DICOM header information on images. This could result is an inaccurate date recorded on the images.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Mfg. Lot or Serial # System ID To be provided GON4833100 01264 860536MR3T KZTBL1700064TJ 860871MR3T KZTBL1800069TJ RADNETMR1700 KZTBL1800022TJ 508872PMR 01302 RADNETMR1708 KZTBL1900130TJ 301698MR1 KZTBL1900071TJ 704333PIO KZTBL1900110TJ 631751MR KZTBL1900107TJ RADNETMR1707 KZTBL1900121TJ 803936LMCMR2 033-4-1532 281HMI750W To be provided 83199160995 To be provided M24313812 To be provided M2850013 To be provided TP0013MR01 To be provided PNUE04MR02 To be provided TP0008MR07 To be provided 083027823408018 To be provided 083027823888617 To be provided GON4826677 To be provided GON4762453 To be provided B5142305 To be provided ZC0002 To be provided ZC0034 To be provided ZC0001 To be provided ZC0306 To be provided GON4795546 To be provided 222768MR01 To be provided 34340MRS02 To be provided 10627MRS02
- Product
- SIGNA Pioneer Nuclear Magnetic Resonance Imaging System - Product Usage: is a whole body magnetic resonance scanner designed to support high…
- Recalling firm
- GE Healthcare, LLC
- Quantity
- 62 (30 US, 32 OUS) total devices
- Distribution
- Worldwide distribution - US nationwide distribution and countries of Australia, China, France, Germany, India, Italy, Japan, Netherlands, Oman, Spain, Sweden, Switzerland, Turkey, and United Kingdom.
- Recall date
- February 5, 2020
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)