Solitaire-C Cervical Spacer System a) 14-531593 recall
Solitaire-C Cervical Spacer System a) 14-531593 Drill/Awl Sleeve b) 14-531594 Spring-Loaded Drill/Awl Sleeve Intended to stabilize cervical spinal…
Why was Solitaire-C Cervical Spacer System a) 14-531593 recalled?
A design deficiency was discovered whereby the Drill/Awl Sleeve and Spring-Loaded Drill/Awl Sleeve instruments (the Sleeves) allow the mating Drill and Awl instruments to penetrate 1.0 mm beyond their intended depths during screw hole preparation.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Part number 14-531593 Part number 14-531594 Catalog number Lot Number 14-531593 N23597 14-531594 N23599
- Product
- Solitaire-C Cervical Spacer System a) 14-531593 Drill/Awl Sleeve b) 14-531594 Spring-Loaded Drill/Awl Sleeve Intended to stabilize cervical spinal…
- Recalling firm
- Ebi, Llc
- Quantity
- 193 units
- Distribution
- US Nationwide Distribution.
- Recall date
- December 31, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)