Model: X-ray TV Sysstem Sonialvision Safire17 recall
MODEL: X-RAY TV SYSSTEM SONIALVISION safire17
Why was Model: X-ray TV Sysstem Sonialvision Safire17 recalled?
It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model: SonialVision safire: Generator Number: D150-40 Serial Numbers: 3M5249D41006 3M5249D2A004 3M5249D34001 3M5249D37006 3M5249D32012 3M5249D35002 3M5249D1C013 3M5249D31009 3M5249D29002 3M5249D33003 3M5249D2C004 3M5249D22013
- Product
- MODEL: X-RAY TV SYSSTEM SONIALVISION safire17
- Recalling firm
- Shimadzu Medical Systems
- Quantity
- 12 systems
- Distribution
- U.S.: AK, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, and WV. O.U.S.: N/A
- Recall date
- April 13, 2022
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)