Model: X-ray R/f System Fluorospeed X1 recall
MODEL: X-RAY R/F SYSTEM FLUOROspeed X1
Why was Model: X-ray R/f System Fluorospeed X1 recalled?
It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model: FLUOROspeed X1 Generator Number: D150-40 Serial Numbers: MQ493E19C001 MQ493E3B8002 MQ493E3B9001 MQ493E1B4001 MQ493E1A1003 MQ493E1A3001 MQ493E1B5001 MQ493E3B8001 MQ493E1A5001 MQ493E1A1005 MQ493E1A2001 MQ493E1A4001 MQ493E1B6002 MQ493E1A8001 MQ493E1B4002 MQ493E1A4003 MQ493E1A4002 MQ493E1B6001 MQ493E1A2002 MQ493E1AC001 MQ493E1A1002 MQ493E19C002 MQ493E1A3003 MQ493E1A1004 MQ493E1A3002
- Product
- MODEL: X-RAY R/F SYSTEM FLUOROspeed X1
- Recalling firm
- Shimadzu Medical Systems
- Quantity
- 25 systems
- Distribution
- U.S.: AK, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, and WV. O.U.S.: N/A
- Recall date
- April 13, 2022
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)