Healon GV PRO recall
Healon GV PRO 0.85 mL, Model (U.S.): 10-2400-14, Model (O.U.S.): 10-2705-14, 10-2715-14, 10-2725-14, and 10-2755-14
Why was Healon GV PRO recalled?
Ophthalmic viscosurgical device may be difficult to remove from the eye, leading to increased postoperative intraocular pressure requiring additional intervention. Also, reports of clogging of phacoemulsion equipment tubing, which may lead to delay in the procedure or ocular injury.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model: 10-2400-14, Lots: UE31164, UE31214, UE31440, UE31478, UE31521, UE31560. Model: 10-2705-14, Lots: UE31098, UE31204, UE31283, UE31306, UE31364, UE31409, UE31467, UE31476, UE31507, UE31519, UE31588, UE31408, UE31439, UE31559, UE31587.
- Product
- Healon GV PRO 0.85 mL, Model (U.S.): 10-2400-14, Model (O.U.S.): 10-2705-14, 10-2715-14, 10-2725-14, and 10-2755-14
- Recalling firm
- Johnson & Johnson Surgical Vision Inc
- Quantity
- N/A
- Distribution
- U.S.: CA, PA, NJ, GA, IL, MO, CO, TX, FL, AK, NY, HI, UT, LA, OH, IN, MA, NC, AZ, MI, NH, AL, TN, WA, MD, OR, VA, MS, MN, KY, AR, NV, WI, WV, SC, DE, ME, CT, ID, NM, VT, IA, OK, MT, NE, DC. OUS: DE, GB, DK, IT, NL, TR, AT, CH, FR, GR, ES, PT, FO, BE, GP, ZA, IE, HR, HU, IS, NO, SE, FI
- Recall date
- February 5, 2020
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)