FDAClass IIFebruary 5, 2020

Atellica CH 930 Analyzer recall

Atellica CH 930 Analyzer, Siemens Material Number 11067000, Software Versions V1.19.2 or 1.20.0 The Atellica CH 930 Analyzer is an automated, clinical…

Why was Atellica CH 930 Analyzer recalled?

On the Atellica CH 930 Analyzer, when scanning barcodes at the module console for Integrated Multisensor Technology (IMT) system fluids (Standard A (Std A) and Standard B (Std B)) for Sodium (Na), Potassium (K), and Chloride (Cl), the system will translate the decimal point separators to commas. This will result in invalid concentration values for the fluids, therefore the system will utilize default nominal values instead of lot specific IMT fluid concentration values for calculation of the calibrations and results for the Na, K, and Cl methods.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI: 00630414002163, Software Versions V1.19.2 or 1.20.0 when used with languages (countries) that use a comma delimiter instead of a decimal separator (i.e. 1,02 vs 1.02), AND when scanning barcodes at the module console.

Product
Atellica CH 930 Analyzer, Siemens Material Number 11067000, Software Versions V1.19.2 or 1.20.0 The Atellica CH 930 Analyzer is an automated, clinical…
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
1528
Distribution
The products were distributed to the following foreign countries: Argentina, Belgium, Brazil, Denmark, Finland, France, Germany, Greece, Italy, Latvia, Netherlands, Poland, Portugal, Spain, Sweden, Switzerland, and Vietnam. No affected products distributed in the United States.
Recall date
February 5, 2020
Classification
Class II
Status
Terminated
Agency
FDA (device)