Bausch & Lomb Stellaris PC w/ Laser recall: Software Issue
Bausch & Lomb Stellaris PC w/ Laser, BL14304 The responsible firm name is Bausch & Lomb Incorporated Rochester, NY
Why was Bausch & Lomb Stellaris PC w/ Laser recalled?
A software anomaly was identified where the system may automatically transition from "Standby" to "Ready to Fire" state while in surgical use.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Serial numbers: SPC01790, SPC01791, SPC01792, SPC01793, SPC01794, SPC01795, SPC01796, SPC01797, SPC01798, SPC01799, SPC01800, SPC01801, SPC01802, SPC01803, SPC01806, SPC01807, SPC01808, SPC01809, SPC01810, SPC01811, SPC01899, SPC01900, SPC01901, SPC01902, SPC01903, SPC01909, SPC01910, SPC01911, SPC01920, SPC01921, SPC01922, SPC01923, SPC01927, SPC01928, SPC01929, SPC01930, SPC01931, SPC01932, SPC01939, SPC01943, SPC01944, SPC01945, SPC01946, SPC01947, SPC01948, SPC01950, SPC01951, SPC01952, SPC01953, SPC01954, SPC01955, SPC01956, SPC01957, SPC01958, SPC01961, SPC01963, SPC01968, SPC01969, SPC01975, SPC01980, SPC01981, SPC01982, SPC01983, SPC01995, SPC01996, SPC01997, SPC02004, SPC02005, SPC02006, SPC02007, SPC02008, SPC02009, SPC02010, SPC02011, SPC02012, SPC02013, SPC02014, SPC02020, SPC02029, SPC02030, SPC02037, SPC02043, SPC02051, SPC02052, SPC02053, SPC02056, SPC02057, SPC02059, SPC02061, SPC02067, SPC02068, SPC02073, SPC02074, SPC02075, SPC02076, SPC02077, SPC02099, SPC02108, SPC02109, SPC02110, SPC02111, SPC02112, SPC02113, SPC02114, SPC02115 and SPC02120
- Product
- Bausch & Lomb Stellaris PC w/ Laser, BL14304 The responsible firm name is Bausch & Lomb Incorporated Rochester, NY
- Recalling firm
- Bausch & Lomb Inc
- Quantity
- 106
- Distribution
- Worldwide Distribution - USA Distribution including to the states of AL, CA, CT, FL, GA, HI, IL, IN, KS, MI, MN, MO, MS, NH, NY, OH, OK, PA, TX, WI and WV., and to the countries of Austria, Australia, Belgium, China, Czech Republic, France, Germany, Great Britain, Greece, Hong Kong, Hungary, India, Ireland, Italy, Jordan, Malaysia, Morocco, Netherlands, Norway, Portugal, Singapore, Spain, Sweden, Switzerland, Turkey and the United Arab Emirates.
- Recall date
- December 31, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)