PageWriter TC30 w/o trolley Government Bundle recall
PageWriter TC30 w/o trolley Government Bundle, Product 860355, Software revisions up to and including A.07.05.22 to evaluate the electrocardiogram of…
Why was PageWriter TC30 w/o trolley Government Bundle recalled?
Philips PageWriter TC cardiograph batteries which have reached end of life may overheat, causing the case to melt and/or the device to ignite, which can cause injury to patient or nearby users.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
US31510433 US31630351 US31630352 US31630353 US51510712 US51510713 US61510947 US61630668 US61630669 US81731977 US81731978 US91511381 US91511445 USO1511461
- Product
- PageWriter TC30 w/o trolley Government Bundle, Product 860355, Software revisions up to and including A.07.05.22 to evaluate the electrocardiogram of…
- Recalling firm
- Philips North America, LLC
- Quantity
- 14
- Distribution
- Domestic distribution nationwide. Foreign distribution worldwide.
- Recall date
- February 27, 2019
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)