FDAClass IIJanuary 20, 2021

Discovery 670 DR recall

Discovery 670 DR (Model H3100BT), NM/CT 870 DR (Model 3100Q and H3100AS), Discovery NM670 Pro (Model 3100XB), Discovery 670 ES (Model H3100XF)…

Why was Discovery 670 DR recalled?

GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Discovery 670 DR (Model H3100BT), Serial Numbers: DRGY56033, DRGY56037, DRGY56035, DRGY56030, DRGZ56043, DRGZ56042, DRGY56032, DRGY56031, DRGY56034, DRGY56036 NM/CT 870 DR (Model 3100Q), Serial Numbers: 870Y64066, and (Model H3100AS), Serial Numbers: 870Y64077, 870Y64064, 870Y64081, 870Y64079, 870Y64067, 870Y64074, 870Z64117, 870Y64078, 870Y64080, 870Y64069, 870Y64085, 870Z64121, 870Z64119, 870Z64122, 870Y64056, 870Y64063, 870Z64114, 870Z64115, 870Y64068, 870Y64082, 870Z64116, 870Z64120, 870Y64065, 870Y64084, 870Y64061, 870Y64083, 870Z64113, 870Y64070, 870Z64123, 870Y64058, 870Z64118, 870Y64057 Discovery NM670 Pro (Model 3100XB), Serial Numbers: PRGY52027, PRGY52031, PRGZ52047, PRGZ52046, PRGY52028, PRGZ52048, PRGY52029, PRGY52030 Discovery 670 ES (Model H3100XF), Serial Numbers: ESGZ53005

Product
Discovery 670 DR (Model H3100BT), NM/CT 870 DR (Model 3100Q and H3100AS), Discovery NM670 Pro (Model 3100XB), Discovery 670 ES (Model H3100XF)…
Recalling firm
GE Healthcare, LLC
Quantity
52 untits
Distribution
Worldwide
Recall date
January 20, 2021
Classification
Class II
Status
Terminated
Agency
FDA (device)