FDAClass IIDecember 31, 2014

Lite DM recall: Software Issue

Lite DM (Lite Data Manager) Software Version(s): Versions 2.0.0.19 and lower Product Model #: D00A00224D The Lite DM (Lite Data Manager) is a software…

Why was Lite DM recalled?

HORIBA Medical is informing all HORIBA Medical Lite DM customers using software version 2.0.0.19 and lower of a software bug related to the potential for results received being applied to the wrong order in the Lite DM Worklist. This bug has not been seen in software versions above V2.0.0.19 IMPACT: Potential for results received being applied to the wrong order in the Lite DM Worklist. The is

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot/Serial #(s): N/A Software Version(s): Versions 2.0.0.19 and lower. Expiration Date (s): N/A

Product
Lite DM (Lite Data Manager) Software Version(s): Versions 2.0.0.19 and lower Product Model #: D00A00224D The Lite DM (Lite Data Manager) is a software…
Recalling firm
Horiba Instruments Inc
Quantity
322 (5 in U.S. Military)
Distribution
Nationwide within the U.S.
Recall date
December 31, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)