FDAClass IIDecember 31, 2014

Persona Cemented Tibial Broach Inserter/Extractor. The recall

Persona Cemented Tibial Broach Inserter/Extractor. The Persona Cemented Tibial Broach Inserter/Extractor Handle is a Use sterilizable instrument…

Why was Persona Cemented Tibial Broach Inserter/Extractor. The recalled?

The affected lot was supplied by contract manufacturer and are potentially subject to disassociation of the magnet cover. Zimmer received one complaint indicating the magnet cover of a Persona Cemented Tibial Broach Inserter/Extractor Handle lot 62085420 was missing after broaching the tibia. Postoperative radiographs were inspected and the magnet cover was not identified. The investigation indi

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Item: 42-5399-023-00 Lots: 62072307 62072565 62085420 62192754

Product
Persona Cemented Tibial Broach Inserter/Extractor. The Persona Cemented Tibial Broach Inserter/Extractor Handle is a Use sterilizable instrument…
Recalling firm
Zimmer, Inc.
Quantity
83 units
Distribution
Distributed in the states of MN; NJ; NC; PA; AL; MI; OR; FL; OH; CO; IL and CA, and the countries of Australia, India, UAE, Germany, and Italy.
Recall date
December 31, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls