RayStation/RayPlan versions 4 - 11A recall
RayStation/RayPlan versions 4 - 11A, including service packs. Build Stations: 4.0.0.14, 4.0.3.4, 4.3.0.14, 4.5.1.14, 4.7.2.5, 4.7.3.13, 4.7.4.4…
Why was RayStation/RayPlan versions 4 - 11A recalled?
An issue where the combined density in a dose grid voxel partially covered by the External ROI and also partially covered by an ROI of type Bolus, Support or Fixation may be unexpected.The density in the voxel can be both under- and overestimated
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Product name (build number) UDI-DI RayStation 4.0 (4.0.0.14) to RayStation 5 Service Pack 2 (5.0.2.35) N/A RayStation 5 Service Pack 3 (5.0.3.17) 07350002010020 RayStation 6/RayPlan 2 (6.0.0.24) 07350002010013 RayStation 6/RayPlan 2 Service Pack 1 (6.1.1.2) 07350002010082 RayStation 6/RayPlan 2 Service Pack 2 (6.2.0.7) 07350002010075 RayStation 6/RayPlan 2 Service Pack 3 (6.3.0.6) 07350002010242 RayStation/RayPlan 7 (7.0.0.19) 07350002010068 RayStation/RayPlan 8A (8.0.0.61) 07350002010112 RayStation/RayPlan 8A Service Pack 1 (8.0.1.10) 07350002010136 RayStation/RayPlan 8B (8.1.0.47) 07350002010129 RayStation/RayPlan 8B Service Pack 1 (8.1.1.8) 07350002010204 RayStation/RayPlan 8B Service Pack 2 (8.1.2.5) 07350002010235 RayStation/RayPlan 9A (9.0.0.113) 07350002010174 RayStation/RayPlan 9B (9.1.0.933) 07350002010266 RayStation/RayPlan 9B Service Pack 1 (9.2.0.483) 07350002010297 RayStation/RayPlan 10A (10.0.0.1154) 07350002010303 RayStation/RayPlan 10A Service Pack 1 (10.0.1.52) 07350002010365 RayStation/RayPlan 10B (10.1.0.613) 07350002010310 RayStation 11A (11.0.0.951) 07350002010389 RayStation 11A Service Pack 1 (11.0.1.29) 07350002010433
- Product
- RayStation/RayPlan versions 4 - 11A, including service packs. Build Stations: 4.0.0.14, 4.0.3.4, 4.3.0.14, 4.5.1.14, 4.7.2.5, 4.7.3.13, 4.7.4.4…
- Recalling firm
- RAYSEARCH LABORATORIES AB
- Quantity
- 478 licenses US
- Distribution
- US Nationwide.
- Recall date
- April 6, 2022
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)