TomoHelical and TomoDirect plans in RayStation 6 recall
TomoHelical and TomoDirect plans in RayStation 6, 7, 8A, 8B, 9A, 9B, 10A, 10B and 11A. Build Numbers: 6.1.1.2, 6.2.0.7, 6.3.0.6, 7.0.0.19, 8.0.0.61…
Why was TomoHelical and TomoDirect plans in RayStation 6 recalled?
If the dynamic jaw mode is used and an optimization is continued after changing the dose grid or modifying the target ROI, the positions of the jaws may change unexpectedly for some control points. User must be aware to avoid unnecessary jaw openings during treatment planning
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Name (build number) UDI-DI: RayStation 6/RayPlan 2 Service Pack 1 (6.1.0.26) 07350002010037 RayStation 6/RayPlan 2 Service Pack 1 v.6.1.1 (6.1.1.2) 07350002010082 RayStation 6/RayPlan 2 Service Pack 2 (6.2.0.7) 07350002010075 RayStation 6/RayPlan 2 Service Pack 3 (6.3.0.6) 07350002010242 RayStation/RayPlan 7 (7.0.0.19) 07350002010068 RayStation/RayPlan 8A (8.0.0.61) 07350002010112 RayStation/RayPlan 8A Service Pack 1 (8.0.1.10) 07350002010136 RayStation/RayPlan 8B (8.1.0.47) 07350002010129 RayStation/RayPlan 8B Service Pack 1 (8.1.1.8) 07350002010204 RayStation/RayPlan 8B Service Pack 2 (8.1.2.5) 07350002010235 RayStation 9A (9.0.0.113) 07350002010174 RayStation 9B (9.1.0.933) 07350002010266 RayStation Service Pack 1 (9.2.0.483) 07350002010297 RayStation 10A (10.0.0.1154) 07350002010303 RayStation 10A Service Pack 1 (10.0.1.52) 07350002010365 RayStation 10B (10.1.0.613) 07350002010310 RayStation 11A (11.0.0.951) 07350002010389 RayStation 11A Service Pack 1 (11.0.1.29) 07350002010433
- Product
- TomoHelical and TomoDirect plans in RayStation 6, 7, 8A, 8B, 9A, 9B, 10A, 10B and 11A. Build Numbers: 6.1.1.2, 6.2.0.7, 6.3.0.6, 7.0.0.19, 8.0.0.61…
- Recalling firm
- RAYSEARCH LABORATORIES AB
- Quantity
- 51 (US) Licenses
- Distribution
- US Nationwide distribution in the states of AL, CA, FL, GA, IA, IL, MO, MS, NC, OK, OR, SC, TN, WA, WI.
- Recall date
- April 6, 2022
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)