FDAClass IIJanuary 29, 2020

Instrumentarium Dental Express SOLID STATE X-RAY IMAGER recall: Impurity Out of Spec

Instrumentarium Dental Express SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)

Why was Instrumentarium Dental Express SOLID STATE X-RAY IMAGER recalled?

Slow degradation of plastic insulation used in the PSU DC plug of the DIGORA Optime, Scan eXam, Express, GXPS-500 systems may result in short circuiting which may cause heating and melting. The melting may generate smoke and eventually cause the PSU to fail.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Number: TBD (range from IJ1208217 to IJ1313309) PSU's having nine-digit serial numbers beginning with four digits as follows: 1143XXXXX 1221XXXXX 1224XXXXX and 1229XXXXX

Product
Instrumentarium Dental Express SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)
Recalling firm
Palodex Group Oy
Quantity
4,563 Total Globally (all products)
Distribution
US: WI and PA OUS: ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COSTA RICA, CZECH REPUBLIC, DENMARK, ECUADOR, FINLAND, FRANCE, GERMANY, GREECE, HUNGARY, INDIA, ISRAEL, ITALY, JAPAN, KOREA, REPUBLIC OF, KUWAIT, MALAYSIA, MALTA, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PAKISTAN, POLAND, QATAR, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVENIA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, REPUBLIC OF CHINA, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, VIETNAM
Recall date
January 29, 2020
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls