Stryker Cassette Pump recall
STRYKER CASSETTE PUMP, AHTO Tube Set, labeled sterile. Includes the following: a. Stryker AHTO Tube Set Packaging (Model 0250070600); b. Stryker AHTO Tube Set with Tip Packaging (Model 0250070620);
Why was Stryker Cassette Pump recalled?
Routine bioburden testing of certain lots were found to have levels higher than internal acceptable rates.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
a. Model 0250070600, UDI 07613327061499, Lot No. 17284FG2, 17291FG2, 17300FG2; b. Model 0250070620, UDI 07613327061659, Lot No. 17275FG2, 17279FG2, 17280FG2, 17282FG2, 17283FG2, 17284FG2, 17285FG2, 17286FG2, 17289FG2, 17291FG2, 17292FG2, 17293FG2, 17297FG2, 17298FG2, 17299FG2;
- Product
- STRYKER CASSETTE PUMP, AHTO Tube Set, labeled sterile. Includes the following: a. Stryker AHTO Tube Set Packaging (Model 0250070600); b. Stryker AHTO Tube Set with Tip Packaging (Model 0250070620);
- Recalling firm
- Stryker Corporation
- Quantity
- 60,753 units total
- Distribution
- Distributed domestically to . Distributed internationally to Australia and Mexico.
- Recall date
- March 14, 2018
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)