PTS Diagnostics CardioChek Plus Professional Analyzer recall
PTS Diagnostics CardioChek Plus Professional Analyzer, New analyzer, boxed, with carrying case, Reference Numbers 2700
Why was PTS Diagnostics CardioChek Plus Professional Analyzer recalled?
One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All Serial Numbers shipped during the 11/1/2016 to 10/25/2019 timeframe. UDI: 00381932700016
- Product
- PTS Diagnostics CardioChek Plus Professional Analyzer, New analyzer, boxed, with carrying case, Reference Numbers 2700
- Recalling firm
- Polymer Technology Systems, Inc.
- Quantity
- 44,448 total
- Distribution
- The products were distributed US Nationwide. The products were distributed to the following foreign countries: Australia, Austria, Belgium, Bermuda, Canada, China, Chile, Costa Rica, Cayman Islands, Czech Republic, England, Finland, France, Germany, Greece, Hong Kong, India, Iraq, Italy, Jordan, Lebanon, Lithuania, Mexico, Morocco, Panama, Philippines, Poland, Qatar, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Trinidad, UAE.
- Recall date
- January 29, 2020
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)