FDAClass IIFebruary 7, 2024

Custom procedural convenience kits and trays recall

Custom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled as: a) C-SECTION PACK, kit number FDCS01AA; b) C-SECTION PACK, kit number FDCS01AB; c) C-SECTION STERILE PACK, kit number HGCS68J; d) C-SECTION PACK, kit number HICS50H; e) C-SECTION PPS, kit number LLCI66; f) C-Section drape Pack, kit number RBCS12R; g) C-Section, kit number THCS49B; h) C-SECTION PACK (PS 020632), kit number UICS50AG;

Why was Custom procedural convenience kits and trays recalled?

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

a) FDCS01AA, UDI/DI 00191072179890, Lot Numbers: 708221, exp. 10/23/2023; 646221, exp. 12/20/2023; b) FDCS01AB, UDI/DI 00191072185273, Lot Numbers: 886231, exp. 4/24/2024; c) HGCS68J, UDI/DI none, Lot Numbers: 702221, exp. 10/22/2023; 654221, exp. 12/12/2023; 960231, exp. 2/9/2024; 903231, exp. 4/7/2024; d) HICS50H, UDI/DI none, Lot Numbers: 909221, exp. 4/1/2023; 835221, exp. 6/14/2023; 755221, exp. 9/2/2023; e) LLCI66, UDI/DI 00191072119025, Lot Numbers: 750221, exp. 9/7/2023; 727221, exp. 9/30/2023; 678221, exp. 11/9/2023; 994231, exp. 11/9/2023; 948231, exp. 11/9/2023; 926231, exp. 11/9/2023; f) RBCS12R, UDI/DI 00191072131812, Lot Numbers: 770221, exp. 8/18/2023; 695221, exp. 11/1/2023; g) THCS49B, UDI/DI 00191072153357, Lot Numbers: 770221, exp. 8/18/2023; h) UICS50AG, UDI/DI none, Lot Numbers: 679221, exp. 11/17/2023; 946231, exp. 2/23/2024; 876231, exp. 5/4/2024;

Product
Custom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled as: a) C-SECTION PACK, kit number FDCS01AA; b) C-SECTION PACK, kit number FDCS01AB; c) C-SECTION STERILE PACK, kit number HGCS68J; d) C-SECTION PACK, kit number HICS50H; e) C-SECTION PPS, kit number LLCI66; f) C-Section drape Pack, kit number RBCS12R; g) C-Section, kit number THCS49B; h) C-SECTION PACK (PS 020632), kit number UICS50AG;
Recalling firm
American Contract Systems, Inc.
Quantity
1818 units
Distribution
US Nationwide distribution.
Recall date
February 7, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls