FDAClass IIMarch 7, 2018

ST Aia-pack¿ Prog III recall

ST AIA-PACK¿ PROG III, Progesterone III Assay, Part Number: 025240

Why was ST Aia-pack¿ Prog III recalled?

Falsely elevated progesterone assay values will occur when using certain progesterone assay part numbers due to cross reactivity of DHEA-S, a metabolite of the DHEA (dehydroepiandrosterone) supplement.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All lots

Product
ST AIA-PACK¿ PROG III, Progesterone III Assay, Part Number: 025240
Recalling firm
Tosoh Bioscience Inc
Quantity
3322
Distribution
Distributed in 39 states: AK, AL, AR, AZ, CA, CO, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, and WI.
Recall date
March 7, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls