FDAClass IIMarch 2, 2016

This reagent is for in vitro diagnostic use. Pax-5 recall

This reagent is for in vitro diagnostic use. Pax-5 (1EW) monoclonal antibody is intended to be used for the qualitative identification by light…

Why was This reagent is for in vitro diagnostic use. Pax-5 recalled?

The Bond" Polymer Refine Detection and Novolink" Polymer Detection System may not provide adequate staining when using a detection protocol where one of its components - Peroxide Block / Peroxidase Block is applied after the primary marker incubation as specified in the Instructions for Use Following the Instructions for Use.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Product Code/Name: PA0552 (All Lots) Bond Ready-to-Use Primary Antibody PAX-5 (1EW). Detection Kit: DS9800 BondTM Polymer Refine Detection Kit. Affected Detection Kit Lots: 42314, 42388, 42474, 42486, 42499, 42555, 42570, 42579, 42627, 42645, 42683, 42780, 42802, 42823, 42852, 43050.

Product
This reagent is for in vitro diagnostic use. Pax-5 (1EW) monoclonal antibody is intended to be used for the qualitative identification by light…
Recalling firm
Leica Microsystems, Inc.
Quantity
1700 units
Distribution
US Consignees incluude: AR ,AZ ,CA ,CO ,CT ,DC ,DE ,FL ,GA ,HI ,IA ,IL ,KY ,LA ,MA ,MD ,MI ,MN ,MO ,MT ,NC ,NH ,NJ ,NY ,OH ,OK ,PA ,PR ,SC ,TN ,TX ,UT ,VA ,VT ,WA ,WI and WV. Foreign Consignees include:Argentina , Australia , Brazil , Bulgaria , Chile , China , Ecuador ,Egypt , El Salvador , Finland , France , Georgia , Germany , Greece , India , Italy , Japan , Jordan , Kuwait , Lebanon , Malaysia , Philippines , Poland , Portugal , Romania , Russian Fed. , Singapore , South Korea , Spain , Sweden , Switzerland , Thailand , Tunisia , Turkey , United Kingdom and United Arab Emirates.
Recall date
March 2, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)