FDAClass IIFebruary 20, 2019

MOSAIQ¿ Oncology Information System and Sequencer recall

MOSAIQ¿ Oncology Information System and Sequencer MOSAIQ is an oncology information system used to manage workflows for treatment planning and…

Why was MOSAIQ¿ Oncology Information System and Sequencer recalled?

There is a potential that Wedge IDs were not included in the DICOM RT PLAN sent from the TPS, which will cause the field in MOSAIQ to be created with no wedge. Treatment of a field without the planned wedge will receive more dose than the treatment plan indicates.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

MOSAIQ Radiation Oncology version 1.0 and higher. Multi-ACCESS version 6.1 and higher. UDI 858164002084 858164002091 858164002107 858164002220 858164002237

Product
MOSAIQ¿ Oncology Information System and Sequencer MOSAIQ is an oncology information system used to manage workflows for treatment planning and…
Recalling firm
Elekta, Inc.
Quantity
3,950 units total (1,747 US)
Distribution
U.S. Nationwide distribution to the following states AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the District of Columbia and Puerto Rico.
Recall date
February 20, 2019
Classification
Class II
Status
Ongoing
Agency
FDA (device)