Titanium Single Lumen Low-Profile port recall: Labeling Error
Titanium Single Lumen Low-Profile port, with Pre-Attached open-Ended Silicone 6.6Fr Catheter. Product Code 0602180 Usage: The 6.6 Fr Open-Ended…
Why was Titanium Single Lumen Low-Profile port recalled?
Bard Access Systems is recalling various Low Profile Port with 6.6 Fr Single Lumen Broviac Catheter because a small number of Silicone catheters in the may be undersized to 5.3 Fr instead of 6.6 Fr as stated in the labeling for the device.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot REXC0467
- Product
- Titanium Single Lumen Low-Profile port, with Pre-Attached open-Ended Silicone 6.6Fr Catheter. Product Code 0602180 Usage: The 6.6 Fr Open-Ended…
- Recalling firm
- Bard Access Systems
- Quantity
- 200
- Distribution
- Worldwide Distribution - US nationwide, Belgium, and Canada.
- Recall date
- January 29, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)