FDAClass IIJanuary 31, 2024

TactiFlex Ablation Sensor Enabled Ablation Catheter recall

TactiFlex Ablation Sensor Enabled Ablation Catheter, Models A-TFSE-FJ, 7.5 F shaft and an 8F distal section, Sterile.

Why was TactiFlex Ablation Sensor Enabled Ablation Catheter recalled?

When the catheters are used with the EnSiteX EP System, the system does not correctly recognize the catheter as the appropriate product model and instead reads and displays it as a different model. This results in some features specific to the catheter do not appear on the display and the catheter image may also appear inverted.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot numbers 10073391, 10075618, 10080798, 10086492, 10087953, 10089624, 10092157, 10093066, and 10095363, Exp. 11/30/2024, GTIN 05415067034618.

Product
TactiFlex Ablation Sensor Enabled Ablation Catheter, Models A-TFSE-FJ, 7.5 F shaft and an 8F distal section, Sterile.
Recalling firm
St. Jude Medical, Atrial Fibrillation Division, Inc.
Quantity
483 units
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CO, DC, DE, FL, GA, IA, IL, KS, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NV, NY, OH, PA, SC, TN, TX, UT, VA, WA, and WV. The country of Canada.
Recall date
January 31, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls