FDAClass IIMarch 7, 2018

OTTO BOCK Kenevo knee joints recall: Labeling Error

OTTO BOCK Kenevo knee joints Model 3C60

Why was OTTO BOCK Kenevo knee joints recalled?

Otto Bock Healthcare Products GmbH has identified a design issue through extended durability testing where the pylon clamping mechanism has the potential to cause damage to the pylon. When used longer than 1.45 million cycles (which represents approximately 2.3 years of use of the knee joint under heavy usage conditions), AND the pylon clamp is overtightened according to the labeled torque specification, AND the patient weighs at most 125kg it may be remote that the pylon breaks, and causing the patient to fall. There have been no failures or complaints reported directly attributed to the failure mode collapsed pylon. In the US Market and in Canada, labeling requires a condition based service at three years. In the rest of the world, a mandatory two year service interval is specified. Only the devices in distribution in the United States and Canada are affected by this action.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Numbers: 201520007, 201520010, 201520012, 201520008, 201520009, 201522040, 201520002, 201520011, 201520006, 201522046, 201522043, 201520028, 201528134, 201528139, 201520013, 201522045, 201525069, 201525109, 201524051, 201524050, 201549087, 201550012, 201550020, 201550022, 201550017, 201550028, 201550027, 201550016, 201550036, 201550014, 201551005, 201550025, 201551028, 201551030, 201550040, 201551032, 201602031, 201602030, 201603014, 201603016, 201603023, 201602034, 201602036, 201602032, 201551034, 201603029, 201603027, 201603028, 201610008, 201601018, 201610010, 201610015, 201610019, 201610022, 201610020, 201611001, 201611010, 201520027, 201610030, 201610035, 201611012, 201603025, 201611002, 201611018, 201612018, 201612020, 201613021, 201613019, 201615008, 201618017, 201616010, 201613008, 201616004, 201616005, 201616012, 201618023, 201618021, 201618018, 201618019, 201618024, 201617031, and 201618025.

Product
OTTO BOCK Kenevo knee joints Model 3C60
Recalling firm
Otto Bock Healthcare Product
Quantity
75 units
Distribution
AK, AR, AZ, CA, CT, DC, GA, IL, IN, KY, LA, MD, MI, MN, NC, ND, NJ, MN, NY, OH, OK, OR, PA, PR, TX, VA, and WA AU, BE, DE, FR, GB, IT, KW, LU, NL, NO, AT, SA, SE, CH, ES, ZA, and HU
Recall date
March 7, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)