SmartPath to dStream for 1.5T recall
SmartPath to dStream for 1.5T, Model No. 781260, 782112
Why was SmartPath to dStream for 1.5T recalled?
An incorrect electrical shock warning label may have been placed on the Liquid Cooling Cabinet (LCC) of affected MR systems. The caution symbol on the incorrect label has a white background instead of yellow, and the label is missing the high voltage warning symbol. This presents a potential risk of electrical shock for service personnel.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model No. 781260 UDI-DI 00884838095076 Serial Numbers 21020 9024 8657 21200 21223 18948 32131; Model No. 782112 UDI-DI 00884838098886 Serial Numbers 30089 20204
- Product
- SmartPath to dStream for 1.5T, Model No. 781260, 782112
- Recalling firm
- Philips North America
- Quantity
- 7 US; 16 ROW total
- Distribution
- Worldwide - US Nationwide distribution in the states of AR, CO, IL, MD, MS, NE, TX and the countries of Brazil, Canada, China, Germany, Greece, India, Mexico, and Saudi Arabia.
- Recall date
- January 31, 2024
- Classification
- Class III
- Status
- Ongoing
- Agency
- FDA (device)