FDAClass IIMarch 7, 2018

NxStage Dialysate Sack: 1) recall: Endotoxin

NxStage Dialysate Sack: 1) REF SAK-301 UDI-M535SAK3010, 2) REF SAK-302 UDI-M535SAK3020, 3) REF SAK-303 UDI-M535SAK3030, 4) REF SAK-304…

Why was NxStage Dialysate Sack: 1) recalled?

May contain endotoxin levels which have been confirmed to exceed the ANSI/AAMI quality standard dialysis fluid of 0.5 EU/ml when used to prepare product.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

SAK-301- Lot # 70879245, Exp Date 7/11/2019 SAK-302- Lot # 71079056, Exp Date 7/30/2019 SAK-303- Lot #'s 70879181 & 70879224, SAK-304- Lot #'s 70779236, 70879045, 70879105, 70879220 & 70979066, Exp Date 7/8/2019 SAK-306- Lot # 70979110, Exp Date 7/24/2019 SAK-307- Lot #'s 70879187 & 71079144, Exp Date 8/27/2019 SAK-402- Lot # 70879235, Exp Date 7/11/2019 SAK-405- Lot #'s 70779202 & 70879233, Exp Date 7/9/2019 SAK-406- Lot #'s 70979253, Exp Date 7/29/2019 SAK-407- Lot #'s 70879236 , Exp Date 7/11/2019

Product
NxStage Dialysate Sack: 1) REF SAK-301 UDI-M535SAK3010, 2) REF SAK-302 UDI-M535SAK3020, 3) REF SAK-303 UDI-M535SAK3030, 4) REF SAK-304…
Recalling firm
NxStage Medical, Inc.
Quantity
16,039 cases/2 units
Distribution
Worldwide Distribution - USA (nationwide) and to the countries of : United Kingdom & Spain
Recall date
March 7, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls