FDAClass IIDecember 28, 2016

Oscor Breezeway 10f S60.8cm D65.6cm Firm TIP recall

OSCOR BREEZEWAY 10F S60.8CM D65.6CM FIRM TIP Catalog # AB1061FT Is designed to facilitate the introduction of catheters to any of the heart chambers…

Why was Oscor Breezeway 10f S60.8cm D65.6cm Firm TIP recalled?

Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot #'s: C1-08260 Exp. October 6, 2016 and C1-08283 October 14, 2016

Product
OSCOR BREEZEWAY 10F S60.8CM D65.6CM FIRM TIP Catalog # AB1061FT Is designed to facilitate the introduction of catheters to any of the heart chambers…
Recalling firm
Oscor, Inc.
Quantity
255
Distribution
Worldwide Distribution - US including CA, NJ, MD, TX, FL, NC, MO, and Puerto Rico. Internationally to AUSTRIA, BOLIVIA, CANADA, CHILE, CYPRUS, NETHERLANS, EGYPT, GERMANY, HONG KONG, ISRAEL, ITALY, SWITZERLAND, and VIETNAM
Recall date
December 28, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls