FDAClass IIFebruary 20, 2013

Mizuho Orthopedic Table recall

Mizuho Orthopedic Table, model 5855, sub assembly 5855-901 Orthopedic Table Spar.

Why was Mizuho Orthopedic Table recalled?

Possible instability of Orthopedic Table Top. If the epoxy bond fails, the aluminum plug may migrate out of the carbon fiber tube. If this occurs, the Orthopedic Table Spar may need to be replaced to ensure stability of the Orthopedic Table Top.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Table model 5855, sub assembly 5855-901 Orthopedic Table Spar, manufactured between 1996 and 2007.

Product
Mizuho Orthopedic Table, model 5855, sub assembly 5855-901 Orthopedic Table Spar.
Recalling firm
Mizuho OSI
Quantity
869 devices
Distribution
USA Distribution -- Worldwide Distribution including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and WY., and the countries of Australia, Canada, Egypt, Finland, France, Germany, China, Israel, Italy, Japan, Netherlands, New Zealand, P.R. China, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Taiwan R.O.C., UAE and UK.
Recall date
February 20, 2013
Classification
Class II
Status
Terminated
Agency
FDA (device)