FDAClass IIFebruary 24, 2016

EVOLIS Microplate System recall

EVOLIS Microplate System, Catalog # 89601. Part number 89788 for the EVOLIS Operator's Manual. in vitro diagnostic Product Usage: EVOLIS Microplate…

Why was EVOLIS Microplate System recalled?

The error "Washer Reagent Clean Fluid" level low alarmed, which prompted the customer to open the system drawer. When the system drawer was opened the customer noted the plate transport was still running and as a consequence the plate was pushed off the plate transporter.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

The EVOLIS Microplate System instrument is not lot controlled, however for traceability purposes each instrument is assigned a unique serial number. SERIAL NUMBERS: 9163700126, 9163700175, 9163700187, 9163700196, 9163700208, 9163700210, 9163700231, 9163700266, 9163700278, 9163700320, 9163700328, 9163700336, 9163700337, 9163700349, 9163700354, 9163700368, 9163700402, 9163700431, 9163700449, 9163700478, 9163700480, 9163700481, 9163700495, 9163700498, 9163700499, 9163700501, 9163700510, 9163700527, 9163700530, 9163700531, 9163700544, 9163700556, 9163700558, 9163700559, 9163700560, 9163700567, 9163700585, 9163700587, 9163700594, 9163700595, 9163700623, 9163700624, 9163700625, 9163700629, 9163700656, 9163700695, 9163700696, 9163700722, 9163700743, 9163700744, 9163700745, 9163700764, 9163700765, 9163700772, 9163700775, 9163700776, 9163700810, 9163700821, 9163700822, 9163700823, 9163700835, 9163700836, 9163700843, 9163700844, 9163700846, 9163700850, 9163700851, 9163700853, 9163700855, 9163700857, 9163700862, 9163700863, 9163700864, 9163700868, 9163700873, 9163700874, 9163700875, 9163700877, 9163700879, 9163700883, 9163700884, 9163700885, 9163700887, 9163700888, 9163700889, 9163700890, 9163700891, 9163700892, 9163700893, 9163700895, 9163700898, 9163700899, 9163700901, 9163700902, 9163700906, 9163700907, 9163700909, 9163700910, 9163700911, 9163700912, 9163700913, 9163700914, 9163700916, 9163700917, 9163700918, 9163700919, 9163700920, 9163700921, 9163700922, 9163700926, 9163700927, 9163700928, 9163700929, 9163700930, 9163700931, 9163700932, 9163700933, 9163700934, 9163700935, 9163700936, 9163700937, 9163700938, 9163700939, 9163700941, 9163700942, 9163700943, 9163700945, 9163700946, 9163700947, 9163700948, 9163700949, 9163700950, 9163700952, 9163700956, 9163700957, 9163700959, 9163700960, 9163700961, 9163700962, 9163700964, 9163700965, 9163700966, 9163700967, 9163700968, 9163700972, 9163700973, 9163700974, 9163700975, 9163700977, 9163700978, 9163700979, 9163700980, 9163700981, 9163700982, 9163700984, 9163700985, 9163700987, 9163700988, 9163700989, 9163700991, 9163700992, 9163700993, 9163700996, 9163700997, 9163700998, 9163700999, 9163701000, 9163701001, 9163701002, 9163701003, 9163701004, 9163701005, 9163701006, 9163701007, 9163701008, 9163701009, 9163701010, 9163701011, 9163701012, 9163701013, 9163701014, 9163701015, 9163701017, 9163701018, 9163701019, 9163701020, 9163701021, 9163701022, 9163701023, 9163701024, 9163701025, 9163701026, 9163701027, 9163701028, 9163701029, 9163701030, 9163701031, 9163701032, 9163701033, 9163701034, 9163701035, 9163701036, 9163701037, 9163701038, 9163701039, 9163701040, 9163701041, 9163701042, 9163701043, 9163701044, 9163701045, 9163701046, 9163701047, 9163701048, 9163701049, 9163701051, 9163701052, 9163701053, 9163701054, 9163701056, 9163701057, 9163701058, 9163701059, 9163701060, 9163701061, 9163701063, 9163700168R, 9163700172R, 9163700176R, 9163700202R, 9163700206R, 9163700209R, 9163700215R, 9163700221R, 9163700234R, 9163700236R, 9163700242R, 9163700246R, 9163700247R, 9163700276R, 9163700281R, 9163700282R, 9163700293R, 9163700294R, 9163700296R, 9163700326R, 9163700327R, 9163700330R, 9163700355R, 9163700359R, 9163700397R, 9163700398R, 9163700400R, 9163700432R, 9163700479R, 9163700494R, 9163700562R, 9163700563R, 9163700564R, 9163700565R, 9163700573R, 9163700617R, 9163700811R, 9163700834R, 9163700837R, 9163700878R, 9163700881R, 9163700886R, 9163700904R, 9163700908R, 9163700915R, 9163700925R, 9163700951R, 9163700958R, 9163700969R, 9163700971R, 9163700986R, and 916370726R.

Product
EVOLIS Microplate System, Catalog # 89601. Part number 89788 for the EVOLIS Operator's Manual. in vitro diagnostic Product Usage: EVOLIS Microplate…
Recalling firm
Bio-Rad Laboratories, Inc.
Quantity
278 units
Distribution
US Nationwide Distribution including Puerto Rico.
Recall date
February 24, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls