FDAClass IFebruary 20, 2019

Medtronic implantable pulse generator: ADAPTA recall

Medtronic implantable pulse generator: ADAPTA, Dual chamber pacemaker (DDD): (a) Model Number ADD01 (b) Model Number ADDR01 (c) Model Number ADDR03…

Why was Medtronic implantable pulse generator: ADAPTA recalled?

A subset of Medtronic dual chamber pacemakers distributed worldwide between 10 March 2017 and 7 January 2019 under the brand names Adapta, Versa, and Sensia when programmed to a dual chamber mode with atrial-sensing, may experience a circuit error that affects device functionality

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

ADAPTA, Dual chamber pacemaker (DDD): (a) Model Number ADD01, GTIN: 00643169708754, All Serial Numbers (b) Model Number ADDR01, GTIN: 00643169708587, 00643169708594, 00643169708600, 0643169708624, 00643169969711, All Serial Numbers (c) Model Number ADDR03, GTIN: 00643169708631, 00643169708648, All Serial Numbers (d) Model Number ADDR06, GTIN: 00643169708655, 00643169708662, All Serial Numbers (e) Model Number ADDRL1, GTIN: 00643169518247, 00643169708679, 00643169708686, 00643169708709, 00643169969803, All Serial Numbers (f) Model Number ADDRS1, GTIN: 00643169708716, 00643169708723, 00643169708730, 00643169708747, 00643169969810, All Serial Numbers (g) Model Number ADVDD01, GTIN: 00643169708761, 00643169708778, 00643169969735, All Serial Numbers

Product
Medtronic implantable pulse generator: ADAPTA, Dual chamber pacemaker (DDD): (a) Model Number ADD01 (b) Model Number ADDR01 (c) Model Number ADDR03…
Recalling firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Quantity
41802 units
Distribution
Wordlwide
Recall date
February 20, 2019
Classification
Class I
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls