FDAClass IIJanuary 15, 2025

Vacuette Safelink recall

VACUETTE SAFELINK, REF: 450210

Why was Vacuette Safelink recalled?

The VACUETTE SAFELINK Holder with male luer lock may leak blood due to the protective rubber sleeve bunching up and not retracting when using the device.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI: 19120049207487, Lot: A240338Q, Expiration Date: 270215

Product
VACUETTE SAFELINK, REF: 450210
Recalling firm
Greiner Bio-One GmbH
Quantity
188,800 pcs
Distribution
US Nationwide distribution in the states of IL, PA.
Recall date
January 15, 2025
Classification
Class II
Status
Ongoing
Agency
FDA (device)