BriteBlade Pro Single-se Fiber Optic Miller 2 recall
BriteBlade Pro Single-se Fiber Optic Miller 2, CE 0120, Do not Re-use, Rx Only, REF, 040-722U, Sterile EO, Qty 10, (01) 15055788722783 The device is…
Why was BriteBlade Pro Single-se Fiber Optic Miller 2 recalled?
The firm has received reports that the spring/washer/bearing components in the block of the laryngoscope blade become loose and separated from the device. If the blade is disengage over the patient's mount it is a possibility that any loose parts could enter the oral cavity which could cause a delay in the procedure while retrieving loose parts.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Part Number/Lot Number 040-722U 181000115 I CCW 040-722U 181001520 I CCW 040-722U 181200006 I CCW 040-722U 181202055 I CCW 040-722U 190100006 I CCW
- Product
- BriteBlade Pro Single-se Fiber Optic Miller 2, CE 0120, Do not Re-use, Rx Only, REF, 040-722U, Sterile EO, Qty 10, (01) 15055788722783 The device is…
- Recalling firm
- Flexicare Medical Ltd.
- Quantity
- 1,024,890 devices
- Distribution
- US: Nationwide OUS: Australia, Canada, China, Czech Republic, Denmark, Finland, Ireland, Italy, Kuwait, Lebanon, Netherlands, Panama, Romania, Russia, South Korea, UAE, UK
- Recall date
- January 22, 2020
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)