FDAClass IIFebruary 20, 2013

Terumo CDI 101 Hematocrit/Oxygen Saturation Monitoring recall: Software Issue

Terumo CDI 101 Hematocrit/Oxygen Saturation Monitoring System. Monitors blood/oxygen saturation, hematocrit and hemoglobin levels.

Why was Terumo CDI 101 Hematocrit/Oxygen Saturation Monitoring recalled?

Inaccurate readings. Software contains unused (turned off) accuracy correction factors for slope (gain) and offset (bias) for Hematocrit/Saturation calculations. The accuracy on any cuvette size did not meet the specifications.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Catalog number: 101 and Serial Numbers: 0503, 1000, 1017, 1070, 1071, 1073, 1074, 1076, 1077, 1080, 1082, 1083, 1107-1110, 1115-1117,1124,1126-1134, 1136, 1137, 1139-1146, 1148-1152, 1154-1156, 1159-1162, 1164-1166, 1168-1177, 1182, 1185-1190, 1192,1193, 1195, 1197, 1198, 1225, 1228, 1229, 1237, 1239-1241,1243-1245,1247, 1249-1326, 1332-1339, 1344-1347, 1349, 1350-1363, 1365-1409, 1411-1436,1438, 1440-1447, 1449-1453, 1455-1463,1465-1482,1484-1492,1494-1552,1555-1565, 1567-1600,1602,1605-1607,1609-1611,1613-1623,1625-1629,1632-1651,1653, 1655-1660, 1662-1664, 1666, 1667,1669-1679,1681-1684,1686,1689-1691, 1693-1710,1712-1715,1717-1750,1752-1757,1759-1771,1773-1785,1787,1788,1790-1809, 1811-1831,1833,1838,1839,1855-1858,1866, and TK001-TK011.

Product
Terumo CDI 101 Hematocrit/Oxygen Saturation Monitoring System. Monitors blood/oxygen saturation, hematocrit and hemoglobin levels.
Recalling firm
Terumo Cardiovascular Systems Corporation
Quantity
622 units
Distribution
Worldwide Distribution.
Recall date
February 20, 2013
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls