ARCHITECT Total T3 Reagent Kit used on ARCHITECT recall
ARCHITECT Total T3 Reagent Kit used on ARCHITECT i1000SR or i2000/i2000SR platforms Product Usage: It is a Chemiluminescent Microparticle Immunoassay…
Why was ARCHITECT Total T3 Reagent Kit used on ARCHITECT recalled?
Samples tested using ARCHITECT Free T3 or ARCHITECT Total T3 assays may show depressed results due to a reagent carryover when testing on board with specific assays on the ARCHITECT i1000SR and i2000/i2999SR platforms.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
List number (Kit Configuration): 7K64 20 (4 x 100 test/kit), 7K64 25 (1 x 100 tests/kit), 7K64 30 (4 x 500 tests/kit), 7K64 35 (1 x 500 tests/kit), 7K64 27 (1 x 100 tests/kit), 7K64 32 (4 x 500 tests/kit), 7K64 37 (1 x 500 tests/kit). All lot codes for each of these list numbers are included in the recall.
- Product
- ARCHITECT Total T3 Reagent Kit used on ARCHITECT i1000SR or i2000/i2000SR platforms Product Usage: It is a Chemiluminescent Microparticle Immunoassay…
- Recalling firm
- Abbott Ireland Diagnostics Division
- Quantity
- 191,410 units total
- Distribution
- Worldwide Distribution - US Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NJ, NV, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, PR. International distribution to AFGHANISTAN, ALBANIA, ALGERIA, ANGOLA, ARMENIA, AUSTRIA, AZERBAIJAN, BAHAMAS, BAHRAIN, BANGLADESH, BELARUS, BELGIUM, BOSNIA AND HERZEGOVI, BRUNEI, BULGARIA, BURKINA FASO, CAMEROON, CANADA, CAYMAN ISLANDS, CHILE, CROATIA, CYPRUS, CZECH REPUBLIC, DEM. REP. OF THE CON, DENMARK, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GAZA & JERICHO, GEORGIA, GERMANY, GHANA, GREECE, HONDURAS, HUNGARY, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, IVORY COAST, JAMAICA, JORDAN, KAZAKHSTAN, KENYA, KUWAIT, KYRGYZSTAN, LATVIA, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MALAWI, MALAYSIA, MALDIVES, MALI, MAURITANIA, MAURITIUS, MOLDOVA, MONTENEGRO, MOROCCO, MOZAMBIQUE, NEPAL, NETHERLANDS, NICARAGUA, NIGERIA, NORWAY, OMAN, PAKISTAN, PANAMA, PARAGUAY, PERU, POLAND, PORTUGAL, QATAR, REPUBLIC OF YEMEN, ROMANIA, RUSSIA, RWANDA, SAUDI ARABIA, SENEGAL, SERBIA, SINGAPORE, SLOVAKIA,SLOVENIA, SOUTH AFRICA, SPAIN, SRI LANKA, ST. LUCIA, ST. VINCENT, SWEDEN, SWITZERLAND, TANZANIA, THE GAMBIA,TRINIDAD & TOBAGO, TUNISIA, TURKEY, UAE, UGANDA, UKRAINE,UNITED KINGDOM, VIETNAM, ZIMBABWE.
- Recall date
- February 13, 2019
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)