FDAClass IMarch 2, 2016

Shuttle Select Slip-Cath Catheter. Angiographic recall

Shuttle Select Slip-Cath Catheter. Angiographic catheter. For use in angiographic procedures as a conduit for the delivery of contrast media and can…

Why was Shuttle Select Slip-Cath Catheter. Angiographic recalled?

Reports of catheter tip splits and or separation. Loss of device function, medical intervention to retrieve separated segment, or complications resulting from separated tip occluding blood flow to end organs may result. Recall expanded 10/07/2015 to include additional catheters.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All lot numbers of the following Product numbers: SCBR4.5-35-135-P-NS-ANG-SHTL, SCBR4.5-35-135-P-NS-ANG-SHTL-JP, SCBR4.5-35-150-P-NS-0-SHTL, SCBR4.5-35-150-P-NS-ANG-SHTL, SCBR4.5-35-75-P-NS-ANGSHTL

Product
Shuttle Select Slip-Cath Catheter. Angiographic catheter. For use in angiographic procedures as a conduit for the delivery of contrast media and can…
Recalling firm
Cook Inc.
Quantity
9519
Distribution
Worldwide Distribution - US (nationwide including Puerto Rico and US Virgin Islands), Algeria, Antigua, Armenia, Austria, Australia, Bahamas, Belgium, Bangladesh, Bulgaria, Bahrain, Brazil, Brunei Darussalam, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Ghana, Great Britain, Greece Guadeloupe, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kuwait, Kazakhstan, Republic of Korea, Kuwait, Lebanon, Lithuania, Luxemburg, Latvia, Libya, Macau, Malaysia, Malta, Martinique, Mexico, Mongolia, Morocco, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Sri Lanka, Sweden, Switzerland, Spain, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.
Recall date
March 2, 2016
Classification
Class I
Status
Terminated
Agency
FDA (device)