FDAClass IIDecember 24, 2014

PKG, 3MM X 29CM INSERT, DISSECTING FORCEPS, P/N recall

PKG, 3MM X 29CM INSERT, DISSECTING FORCEPS, P/N 0250282110 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and…

Why was PKG, 3MM X 29CM INSERT, DISSECTING FORCEPS, P/N recalled?

Stryker Endoscopy is recalling Laparoscopic manual instruments and accessories because the parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

443338 540049 845866 846795 1040235 1042471 1042780 1043317 1141459 1143139 1144954

Product
PKG, 3MM X 29CM INSERT, DISSECTING FORCEPS, P/N 0250282110 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and…
Recalling firm
Stryker Endoscopy
Quantity
180,573 units total
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Recall date
December 24, 2014
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls