Haylard Closed Suction System for Neonates/Pediatrics recall
Haylard Closed Suction System for Neonates/Pediatrics, 5 Fr, 1. REF 195, Y-Adapter, SAP Finished Product Code 109382303; 2. REF 201, with BALLARD…
Why was Haylard Closed Suction System for Neonates/Pediatrics recalled?
Avanos Medical has received reports stating that the central lumen of some 5 Fr Neonate/Pediatric Closed Suction Catheters may be fully blocked or partially blocked. This may cause the closed suction catheters (CSCs) to inadequately suction secretions and/or saline fluids form the patient's airway to prevent oxygen desaturation.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
REF 195 Lot Code / Expiration Date: M18115T401/2023-APR, M18121T402/2023-MAY, M18128T404/2023-MAY, M18143T401/2023-MAY, M18149T403/2023-MAY, M18156T402/2023-JUN, M18170T402/2023-JUN, M18177T402/2023-JUN, M18192T402/2023-JUL, M18206T401/2023-JUL, M18220T402/2023-AUG, M18247T402/2023-SEP, M18254T405/2023-SEP, M18282T404/2023-OCT, M18311T401/2023-NOV, M18318T402/2023-NOV, M18345T403/2023-DEC, M1834T403/2023-DEC, M18358T403/2023-DEC, M19008T401/2024-JAN, M19058T402/2024-FEB, M19078T403/2024-MAR, M19086T401/2024-MAR, M19098T403/2024-APR, M19134T404/2024-MAY, REF 201 Lot Code / Expiration Date: M18044T405/2023-FEB, M18311T402/2023-NOV, M18358T404/2023-DEC, M19037T402/2024-FEB, M19065T402/2024-MAR, M7172T402/2022-JUN. REF 202 Lot Code / Expiration Date: M18143T403/2023-MAY, M18192T401/2023-JUL, M18254T404/2023-SEP, M18274T401/2023-OCT, M18296T405/2023-OCT, M18318T401/2023-NOV, M19072T401/2024-MAR, M19142T401/2024-MAY, M7346T403/2022-DEC. REF 195-5 Lot Code / Expiration Date: 20000095/2024 - JUL, M18274T402/2023-OCT, M18290T402/2023-OCT, M18296T406/2023-OCT, M18311T403/2023-NOV, M18324T401/2023-NOV, M18324T402/2023-NOV, M18332T401/2023-NOV, M18339T401/2023-DEC, M18345T404/2023-DEC, M18352T402/2023-DEC, M19024T402/2024-JAN, M19030T401/2024-JAN, M19051T401/2024-FEB, M19059T401/2024-FEB, M19065T401/2024-MAR, M19072T402/2024-MAR, M19079T401/2024-MAR, M19085T402/2024-MAR, M19098T404/2024-APR, M19099T401/2024-APR, M19105T403/2024-APR, M19120T403/2024-APR, M19128T401/2024-MAY, M19128T403/2024-MAY, M19135T401/2024-MAY, M19142T402/2024-MAY, M19149T405/2024-MAY, M19162T403/2024-JUN, M19177T402/2024-MAY, M19184T401/2024-JUL.
- Product
- Haylard Closed Suction System for Neonates/Pediatrics, 5 Fr, 1. REF 195, Y-Adapter, SAP Finished Product Code 109382303; 2. REF 201, with BALLARD…
- Recalling firm
- Avanos Medical, Inc.
- Quantity
- 114,300 units (37,880 US, 76,420 OUS)
- Distribution
- Nationwide. Internationally distributed to: Argentina, Australia, Austria, Brazil, Canada, Chile, Columbia, Croatia, Czech Republic, Denmark, France, Germany, Greece, Honduras, India, Ireland, Israel, Italy, Korea, Republic of, Lebanon, Mexico, Nepal, Netherlands, New Zealand, Norway, Peru, Poland, Portugal, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom.
- Recall date
- January 15, 2020
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)