Beckman Coulter DxU 850m/840m Microscopy Analyzers recall
Beckman Coulter DxU 850m/840m Microscopy Analyzers, Part Numbers: C49513, C76947
Why was Beckman Coulter DxU 850m/840m Microscopy Analyzers recalled?
Beckman Coulter has become aware that the cannula, which is found in the Flowcell and the iQClear Rinse Waste Well Assembly, may rust. Rust from the cannula has the potential to generate erroneous results.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
C49513, UDI/DI 15099590748647; C76947, UDI/DI 15099590748654; All serial numbers
- Product
- Beckman Coulter DxU 850m/840m Microscopy Analyzers, Part Numbers: C49513, C76947
- Recalling firm
- Beckman Coulter, Inc.
- Quantity
- 254 units
- Distribution
- US and Australia, Brazil, Canada, China, France, Germany, Ghana, India, Israel, Italy, Lebanon, Lithuania, Mexico, Panama, Qatar, Romania, Saudi Arabia, Spain, Taiwan, Province of China, United Kingdom of Great Britain and Northern Ireland,
- Recall date
- January 31, 2024
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)