PKG, 3MM INSERT, HOOK SCISSORS, 29CM, P/N 0250282049 recall
PKG, 3MM INSERT, HOOK SCISSORS, 29CM, P/N 0250282049 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of…
Why was PKG, 3MM INSERT, HOOK SCISSORS, 29CM, P/N 0250282049 recalled?
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
643889 845498 846278 848131 1041615 1042530 1043321 1043534 1044413 1044450 1046413 1140170 1241796 1344964 1345084
- Product
- PKG, 3MM INSERT, HOOK SCISSORS, 29CM, P/N 0250282049 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of…
- Recalling firm
- Stryker Endoscopy
- Quantity
- 180,573 units total
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
- Recall date
- December 24, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)