Patient transport Shuttle with Trendelenburg quick recall
Patient transport Shuttle with Trendelenburg quick adjustment function (Jupiter, Mars, Titan, TruSystem and Saturn operating table systems). The shuttle is intended for the following applications: 1. Transfer, transport and storage of exchangeable table tops in the Trumpf Medical TruSystem 7500, JUPITER, and SATURN SM operating table systems 2. Patient transport to an operating table top (without a column) from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system 3. Transporting the TruSystem 7500 SM, JUPITER SM or SATURN SM operating table column with table top (without a patient) within the operating theatre.
Why was Patient transport Shuttle with Trendelenburg quick recalled?
Complaints were received of the table top on the transporter (shuttle) tilting down quickly in a maximum Trendelenburg position.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Product: Shuttle 2.7, Material Number: 1254362, Serial Number Range: 100801175 101349420; Product: Shuttle 2.7 360, Material Number: 1459862, Serial Number Range: 100843669-101212847; Product: Shuttle 3.7, Material Number: 1254363, Serial Number Range: 100180383-101290987 and Product: Shuttle 3.7 360, Material Number: 1459863, Serial Number Range: 100787418-101334652.
- Product
- Patient transport Shuttle with Trendelenburg quick adjustment function (Jupiter, Mars, Titan, TruSystem and Saturn operating table systems). The shuttle is intended for the following applications: 1. Transfer, transport and storage of exchangeable table tops in the Trumpf Medical TruSystem 7500, JUPITER, and SATURN SM operating table systems 2. Patient transport to an operating table top (without a column) from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system 3. Transporting the TruSystem 7500 SM, JUPITER SM or SATURN SM operating table column with table top (without a patient) within the operating theatre.
- Recalling firm
- Trumpf Medical Systems, Inc.
- Quantity
- 1284 affected units
- Distribution
- Worldwide Distribution - US including MA and TX; Foreign distribution to Austria, Azerbaijan, Belgium, British Virgin Islands, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Luxemburg, Morocco, Netherlands, New Caledonia, Norway, Poland, Portugal, Qatar, Romania, San Marino, Spain, Sweden and Switzerland.
- Recall date
- February 17, 2016
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)