16600 / Fetzer Surgical De Lee Obstetric Forceps recall
16600 / Fetzer Surgical De Lee Obstetric Forceps, 11" (30cm), Item No./Product Code 46-5034 Fetzer Medical Obstetrical Forceps is a hand-held…
Why was 16600 / Fetzer Surgical De Lee Obstetric Forceps recalled?
The firm had a report of doctors being unable to pull the obstetric forceps apart.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot Number 18800612
- Product
- 16600 / Fetzer Surgical De Lee Obstetric Forceps, 11" (30cm), Item No./Product Code 46-5034 Fetzer Medical Obstetrical Forceps is a hand-held…
- Recalling firm
- Fetzer Medical GmbH & Co. KG
- Quantity
- 10 units
- Distribution
- US Distribution to states of: IL and VA.
- Recall date
- January 30, 2019
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)