FDAClass IIFebruary 6, 2013

Brand Name recall

Brand Name: RX Accunet Embolic Protection System Common Name: RX Accunet EPS. Part Numbers: 101649-45, 1011649-55, 1011649-65, 1011649-75, 1011650-55…

Why was Brand Name recalled?

The recall was initiated because Abbot Vascular has discovered that the identified lots of the RX Accunet Embolic Protection System may exhibit difficult removal of the peel away sheath due to higher than normal wall thickness.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Number: 1081061, 1082561, 1091361, 1080561, 1081761, 1090261, 1091561, 1100461, 1080361, 1081661, 1090161, 1090861, 1092361, 1080861, 1082661, 1100761, 1081761, 1092861, 1082461, 1081161, 1082261, 1083061, 1081961, 1093061.

Product
Brand Name: RX Accunet Embolic Protection System Common Name: RX Accunet EPS. Part Numbers: 101649-45, 1011649-55, 1011649-65, 1011649-75, 1011650-55…
Recalling firm
Abbott Vascular
Quantity
1253 units (962 units in the US & Puerto Rico)
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Austria, Belgium, China, Germany, Spain, France, United Kingdom, Italy, Pakistan, Poland, Russia, Slovakia, Taiwan, and South Africa.
Recall date
February 6, 2013
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls