FDAClass IIJanuary 6, 2021

Roche cobas pro integrated solutions recall

Roche cobas pro integrated solutions (cobas pro ISE, cobas c 503, and cobas e 801 analytical units) Chemistry Analyzer (photometric discrete), for…

Why was Roche cobas pro integrated solutions recalled?

Potential for Changed Configuration Settings on the cobas 8000 modular analyzer series/cobas pro integrated solutions, may cause incorrect results in several affected parameters. In the case of poor sample quality, discrepant results may remain undetected due to the absence of associated data flags.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Numbers: 1925-10; 1931-04; 1931-05;1931-07;1937-10; 1938-01; 1942-10

Product
Roche cobas pro integrated solutions (cobas pro ISE, cobas c 503, and cobas e 801 analytical units) Chemistry Analyzer (photometric discrete), for…
Recalling firm
Roche Diagnostics Operations, Inc.
Quantity
7 units
Distribution
US Nationwide distribution.
Recall date
January 6, 2021
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls