Metzenbaum Scissors 7 recall: Sterility Concern
Metzenbaum Scissors 7, Curved, Sterile The Metzenbaum Scissors 7, Curved Curved are single-use sterile surgical instruments intended for cutting and…
Why was Metzenbaum Scissors 7 recalled?
The Chevron seals of 6x10 Tyvek peel pouch used to package the sterile instruments may separate causing the packages to open and compromise sterility of the instrument.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Catalog # ST5-182; lot# 355
- Product
- Metzenbaum Scissors 7, Curved, Sterile The Metzenbaum Scissors 7, Curved Curved are single-use sterile surgical instruments intended for cutting and…
- Recalling firm
- Integra York PA, Inc
- Quantity
- 50 units
- Distribution
- US Distribution in NC and TX.
- Recall date
- December 21, 2016
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)