Calibration Serum Level 3. CAL 3 recall
Calibration Serum Level 3. CAL 3, REF CAL2351. Human Assayed Multi-Sera Level 2 (HN1530) Human Assayed Multi-Sera Level 3 (HE 1532)
Why was Calibration Serum Level 3. CAL 3 recalled?
There have been transcription errors on the Instructions For Use (IFU) of the Calibration Serum Level 3 (CAL2351 lot 1260UE, 1262UE and 1315UE) for target values of Bilirubin and Triglycerides, on the IFU of Human Assayed Multi-Sera Level 2 (HN1530) for TBIC taget value and on IFU of Human Assayed Multi-Sera Level 3 (HE 1532) for Lipase target value.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
UDI-DI: 05055273200966, lot1260UE, exp. 28 Nov 2024; lot 1262UE, exp. 28 Jan 2025; and lot 1315UE, exp. 28 May 2025
- Product
- Calibration Serum Level 3. CAL 3, REF CAL2351. Human Assayed Multi-Sera Level 2 (HN1530) Human Assayed Multi-Sera Level 3 (HE 1532)
- Recalling firm
- Randox Laboratories Ltd.
- Quantity
- N/A
- Distribution
- US Nationwide distribution.
- Recall date
- January 24, 2024
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)